CERAMENT™| BONE VOID FILLER Mixing Chart - Biomet

Report 6 Downloads 244 Views
If drilling & screw insertion is not required, the wound can be closed anytime after 10 minutes

CERAMENT™ |BONE VOID FILLER Mixing chart

MIX

0

a)

WAIT

30s

INJECT

3 min

WAIT

5 min

MOLD

7 min

DRILLING & SCREW INSERTION

WAIT

9 min

~15 min

a)

b)

b)

WWW.BONESUPPORT.COM PR 0242-03 EN/US

a) Remove plug from large powder syringe b) Attach valve, clear end to the large   powder syringe

Mix for 30 seconds

a) Unscrew plug of liquid filled syringe using provided tool embedded in blister b) Attach liquid syringe to valve (blue end)

Empty liquid into large powder syringe

Remove red plunger stop

Fully retract blue handle and turn blue knob clockwise

Transfer paste into empty delivery syringe

Remove red plunger stop. Wait 3 minutes before injecting from the syringe. Refer to time chart above for further options

For product information, including indications, contraindications, warnings, precautions, and potential side effects see the package insert and www.biomet.com.

©2015 Biomet Biologics • Form No. BMET0910.1 • REV0715 CERAMENT™|BONE VOID FILLER is a trademark of BONESUPPORT AB.