Design of quadrature-compensated double-tuned RF

A study of NOE enhancement in a double-tuned coil at 3T

31P

MRS with respect to

1H

coil quality and size in

C.-H. Choi ([email protected]), Y. Ha, P. Chervakov, N.J. Shah Institute of Neuroscience and Medicine - 4, Forschungszentrum Jülich, Germany

Results

Introduction

 All the 31P data were acquired with an FID sequence with/without NOE (TR/TE=15s/0.35ms, 8 avg., 5kHz BW and for NOE, ten 90ms 180º rectangular pulses with 100ms interval) and processed using jMRUI [5].

 31P signals can be boosted with 1H-decoupled or nuclear Overhauser effect (NOE) [1,2]. Different double-tuned coils have been designed with a minimum 31P coil sensitivity loss [3,4].

 Table 1 summaries the Q-factors of the probe measured on the bench with various forward currents and shows the Q-factor of the coil increases with increasing the forward current on the PIN-diodes. The NOE enhancement also appears to be related to the Q-factor of the 1H coil.

 However, evaluating NOE using double-tuned coils particularly associated with the 1H coil has not been seriously considered. In this study, we designed double-tuned 1H/31P coils and evaluated the NOE effect as a function of the quality/size of the 1H coils.

 Ratio of 31P signal enhancement obtained by larger 1H coil was found to be concomitantly higher (Table 1).  An example of 31P peaks using Coil (A) with 100 mA with and without NOE enhancement is shown in Fig. 2. The Q-factor of 31P coils remains constant and is not affected by changing the level of the forward current of 1H coil.

Methods

Mitglied der Helmholtz-Gemeinschaft

Fig. 1: Double-tuned 1H/31P coil photographs and schematic diagrams – Coil (A) with PINdiodes, Coil (B) without PIN-diodes and Coil (C) with a larger 1H loop.

Table 1. Q-factors (unloaded/loaded/ratio) of the 1H coils and corresponding NOE enhancement values calculated using 31P data with NOE pulses divided by that without NOE pulses.

comparison data

 The double-resonant coils consist of concentrically configured inner (for 31P) and outer (for 1H) loops (Fig. 1). The inner coil includes a trap to isolate coupling at the 1H frequency.  Two sets of experiments - quality and/or size dependence - were performed at 3T using a 30mM KH2PO4 phantom. To vary the quality of the coil, we deliberately inserted two PIN-diodes on the outer 1H loop in (Fig. 1) Coil (A), since the applied forward current adjusts the resistance of the PIN-diodes, and therefore affects the quality of the 1H coil.  In addition, the effect on NOE enhancement using two different sizes of the 1H coil has been compared. The loop size of Coil (C) is ~33% larger than that of Coil (B) in Fig. 1.

Fig. 2: One of the 31P peaks obtained using the Coil (A) 100 mA with/without employing the NOE enhancement technique.

Discussion/Conclusions  The quality/size of the 1H coil may play an important role in contributing to 31P signal enhancement.  When designing double-tuned coils, a 1H coil is often sacrificed to minimise the loss at X-nuclei.  This work provides an excellent reference for coil builders to optimise their design of double-tuned coils for NOE enhancement.

References

1. Lagemaat NMR Biomed. 2016, 2. van Uden ISMRM 2017, 3. Brown Neuroimage 2016, 4. Schaller ISMRM 2016, 5. Stefan Meas. Sci. Tech. 2009.