Precision Balances
FZ FZ-i FX FX-i Series
ISO 9000 CERTIFIED
...Clearly a Better Value http://www.aandd.jp
“High Functionality You Can Afford” The FZ-i/FX-i Series, a new addition to A&D’s precision balances, rivals our GX/GF Series in performance and quality, but offers higher accessibility, bringing to you a cost-effective solution for a diverse array of applications.
Newly developed, Air-pump Controlled Internal Calibration (FZ-i models only) * i
i Patent Pending
For the internal calibration of the FZ-i Series, A&D adopted a direct-driving method that utilizes air pressure for its power source. This new method features a simple structure, high reliability and a fail-safe mechanism – the internal mass returns to the original position even if the calibration is interrupted by power failure, etc.
One Touch Automatic Calibration (FZ-i models only) With just one press of a key, you can ensure the accuracy of your balance.
Compact, B5 Size Footprint By taking advantage of our Compact Super Hybrid Sensor (C-SHS) technology ii , A&D has significantly reduced the size of the FZ-i/FX-i Series – 25% smaller footprint than the GX/GF Series, making it ideal for installation and use in narrow spaces. * ii Patent Pending
Fast, 1 Second Stabilization The C-SHS combines the best of magnetic force restoration and single point parallelogram load sensing technologies. This allows the C-SHS to realize a response speed of just 1 second while maintaining extremely high span stability. The FZ-i/FX-i Series will maximize your productivity and efficiency.
Shock Absorber Function
iii
* iii Patent Pending
The internal shock absorber protects the C-SHS from the adverse effects of vertical and transverse overloading.
Super Hybrid Sensor
VFD Display
Quick USB
LAN-Ethernet Interface
RS-232C Interface
GLP Compliant
Counting Function
Auto Power OFF
Underhook
Rechargeable Battery
Software
Internal Calibration
Percentage Function
Animal Weighing
ON Comparator Function
Auto Power ON
EC Type Approval
(FZ-i)
External Calibration
(FX-i)
Statistical Calculation
RS-232C Interface (standard) Bi-directional communication is possible with a PC, printer or other peripheral devices.
Quick USB Interface (FXi-02) The easy-to-use optional USB interface requires no driver or software installation for transfer iv of weighing data to a PC. * iv Unidirectional communication from a balance to a PC only. Please use the standard RS-232C interface to send commands from a PC to the balance.
LAN-Ethernet Interface (FXi-08) with WinCT-Plus Software WinCT-Plus software makes it possible for the user to easily set up an IP address, subnet mask, etc. The user can also send commands for controlling and acquiring data from multiple balances.
WinCT-Plus
LAN
USB
Ethernet Hub Provided by end-user
FX i -08
FX i -08
FX i -02
FX i -08
Multiple Weighing Units Programmable or standard weighing units of measurement can be selected; g, pcs, %, oz, lb, L oz, ozt, ct, mom, dwt, GN, tl, tol, mes and MLT
GLP/GMP/GCP/ISO Compliant In accordance with GLP and LIMS regulations, the balance ID number and calibration data can be printed out using A&D’s AD-8121B or a PC
Statistical Calculation Function The balance displays and/or output statistical calculation data as No. (number of data), SUM (total), MAX (maximum), MIN (minimum), R [range (maximum – minimum)], AVE (average), SD (standard deviation) and CV (coefficient of variation)
Response Indicator Function Selectable FAST, MID or SLOW response to suit weighing environment conditions.
ACAI – Automatic Counting Accuracy Improvement This counting function recalculates the average unit weight each time a sample is added to eliminate errors caused by variations in unit weight
Built-in Rechargeable Battery (FXi-09) An optional Ni-MH rechargeable battery pack can be fitted into the FZ-i/FX-i Series for use without AC power and for mobile applications (10 hours charging time for 8 hours operation)
Large Breeze Break (FXi-11) The small breeze break comes as standard with all the FZ-i models, FX-120i, FX-200i and FX-300i.
Carat Balance
Specifications
FZ-120i FX-120i 122g
Weighing capacity
FZ-200i FX-200i
FZ-300i FX-300i
FZ-1200i FX-1200i
FZ-2000i FX-2000i
FZ-3000i FX-3000i
220g
320g
1220g
2200g
3200g
Minimum weighing value
0.001g
Counting mode
0.001g
Minimum unit mass
0.01g 0.01g 5, 10, 25, 50 or 100 pieces
Number of samples Percent mode
0.100g
Minimum 100% reference mass
1.00g
0.01%, 0.1%, 1% (depends on the reference mass stored)
Minimum 100% display Repeatability (Standard Deviation) Linearity
0.001g
0.01g
±0.002g
±0.02g Approx. 1 second
Stabilization time (typically set at FAST)
±2ppm /°C
Sensitivity drift (10°C to 30°C)
5°C to 40°C (41°F to 104°F), 85%RH or less (no condensation)
Operating temperature
Vacuum Fluorescent Display (VFD)
Display type
5 times /second, 10 times /second or 20 times /second
Display refresh rate
RS-232C Interface
Standard serial interface
130mm
Weighing pan (diameter) External calibration mass
150mm
100g
200g
300g
1000g
2000g
3000g
50g
100g
200g
500g
1000g
2000g
100g Approx. 11VA (supplied to the AC adaptor)
Power consumption
193 (W) X 262.5 (D) X 85.5 (H) mm
External dimensions
FZ-i Approx. 3.0kg / FX-i Approx. 2.5kg
Net weight
A
168.5 / 6.63
162 / 6.38
61.5 / 2.42
Accessories
194.5 / 7.66 192*
3*
*Underhook
Dimensions with Small Breeze Break (FX i-10)
193 / 7.6
(Comes equipped with FZ,FX-120i /200i /300i , FZ-1200i /2000i /3000i ) 184 / 7.24
184 / 7.24 168.5 / 6.63
115 / 4.53
171.5 / 6.75 175 / 6.89
ø130 / 5.12
81 / 3.19 83.5 / 3.29
115 / 4.53 85.5 / 3.33 85 / 3.35
Static eliminator Electrostatic fieldmeter Weighing data logger Tweezers for calibration weight Remote display Remote controller Compact printer Universal printer Serial/Ethernet converter Quick USB adapter Serial/USB converter RS-232C cable (9P-9P) Carat pan (silver, 2pcs) Carat pan (black, 2pcs) Main unit cover (5pcs)
262.5 / 10.33
A 130 / 5.12 150 / 5.91
Model FZ,FX-120i / 200i / 300i FZ,FX-1200i / 2000i / 3000i
262.5 / 10.33
* FXi-02, FXi-08 and FXi-09 cannot be used at the same time.
Pan size
70.5 / 2.78
Quick USB interface with cable Ethernet interface with WinCT-Plus software Built-in rechargeable battery Small breeze break Large breeze break Animal weighing pan (FZ,FX-1200i / 2000i / 3000i ) Carrying case
AD-1683 AD-1684 AD-1688 AD-1689 AD-8920A AD-8922 AD-8121B AD-8118C AD-8526 AD-8527 AX-USB-9P AX-KO2466-200 AX-CARATPAN-W AX-CARATPAN-B AX-FXi-31
193 / 7.6
External Dimensions (mm/inches)
85.5 / 3.37
Options FXi-02* FXi-08* FXi-09* FXi-10 FXi-11 FXi-12 FXi-15
1000g
Confirm that the adaptor type is correct for the local voltage and power receptacle
AC adaptor
Dimensions with Large Breeze Break (FX i-11) 105/4.13
112/4.41
AD-8121B
294/11.57
315/12.4
199.5/7.85
193/7.6
Ø156/6.14
107/4.21
...Clearly a Better Value 3-23-14 Higashi-lkebukuro, Toshima-ku,Tokyo 170-0013 JAPAN Telephone:[81](3) 5391-6132 Fax:[81](3) 5391-6148 http://www.aandd.jp 1756 Automation Parkway, San Jose, CA 95131 U.S.A. Telephone:[1](408) 263-5333 Fax:[1](408) 263-0119
A&D Australasia ia Pty y Ltd. d. 32 Dew Street, Thebarton, South Australia 5031 AUSTRALIA Telephone:[61](8) 8301-8100 Fax:[61](8) 8352-7409
Unit 24/26 Blacklands Way Abingdon Business Park, Abingdon, Oxon OX14 1DY UNITED KINGDOM Telephone:[44](1235) 550420 Fax:[44](1235) 550485
Vereyskaya str.17, Moscow, 121357 RUSSIA Telephone: [7] (495) 937-33-44 Fax: [7] (495) 937-55-66
509 Udyog Vihar Phase V Gurgaon-122 016, Haryana, INDIA Telephone: [91](124) 471-5555 Fax: [91](124) 471-5599
Große Straße 13 b 22926 Ahrensburg GERMANY Telephone:[49](0) 4102 459230 Fax:[49](0) 4102 459231
A&D D Inst Instruments uments India Pr Private e Limited mited
Manhattan Bldg. 8F, 36-2 Yoido-dong, Youngdeungpo-gu, Seoul, KOREA Telephone:[82](2) 780-4101 Fax:[82](2) 782-4280
*FZi/FXi-ADCC-02-PR2-09a03