TOR 18FG_specSheet

Report 7 Downloads 445 Views
MiRo House Becklands Close Boroughbridge North Yorkshire YO51 9NR United Kingdom

TOR 18FG

FLUOROSCOPY AND FLUOROGRAPHY

A routine test object designed to be used quickly and easily on a regular basis (e.g. weekly/monthly) to provide an ongoing check of imaging performance, particularly those aspects which are most liable to deterioration. After an initial grey-scale check, image quality is measured simply by counting the number of details detected and the number of bar-patterns resolved in the image. An ongoing record of these numbers will reveal any trend towards deterioration in imaging performance. Used for fluoroscopy and fluorography, enables the following checks to be made: Monitor brightness and contrast level adjustment (highlight and lowlight details) Resolution limit (0.5 to 5.0 LP/mm) Low-contrast large-detail detectability (18 details, 8mm diameter, contrast range 0.009 to 0.167 @ 70kVp 1mm Cu) Circular Geometry (Lead Circle)

For more information or to find your regional distributor please visit our website www.leedstestobjects.com

|

Copyright © 2011 Leeds Test Objects Ltd

tel. +44(0)1423 321102 |

SpecSh 174a 16/11/2010

MiRo House Becklands Close Boroughbridge North Yorkshire YO51 9NR United Kingdom

TOR 18FG

FLUOROSCOPY AND FLUOROGRAPHY

TOR 18FG X-RAY IMAGE

For more information or to find your regional distributor please visit our website www.leedstestobjects.com

|

Copyright © 2011 Leeds Test Objects Ltd

tel. +44(0)1423 321102 |

SpecSh 174a 16/11/2010