UK 3042
EU-type examination certificate UK 3042 Revision 1 Issued by:
NMO Notified Body Number 0126 In accordance with the requirements of the Non-automatic Weighing Instruments Regulations 2016 (S.I. 2016 No. 1152) which implement, in the United Kingdom Council Directive 2014/31/EU, this EU-type examination certificate has been issued to:
Avery Weigh-Tronix Foundry Lane Smethwick West Midlands B66 2LP United Kingdom In respect of an Avery Weigh-Tronix instrument designated the ZM505 / ZM510 / ZM605 / ZM615 Series and having the following characteristics: n ≤ 10,000 for Class III instruments with single interval ni ≤ 3,000 for Class III instruments with multi-range or multi-interval operation, per weighing range, with a maximum of three weighing ranges n ≤ 1,000 for Class IIII instruments with single interval The necessary data (principal characteristics, alterations, securing, functioning etc) for identification purposes and conditions (when applicable) are set out in the descriptive annex to this certificate. This revision replaces previous versions of the certificate.
Issue date: Valid until:
13 September 2017 20 July 2026
Grégory Glas Technical Manager For and on behalf of the Head of Technical Services 0135 NMO I Stanton Avenue I Teddington I TW11 OJZ I United Kingdom Tel +44 (0) 20 8943 7272 I Fax +44 (0) 20 8943 7270 I Web www.gov.uk/government/organisations/regulatory-delivery NMO is part of the Regulatory Delivery directorate within the Department for Business, Energy & Industrial Strategy
Descriptive Annex 1
NAME AND TYPE OF INSTRUMENT
This family of instruments utilises the digital indicating devices designated the Avery WeighTronix ZM505 / ZM510 / ZM605 / ZM615 Series (Figures 1 to 8), connected to a weighing platform to form Class III or IIII, mains, DC or battery-powered, self-indicating, NonAutomatic Weighing Instruments. The instruments shall not be used for direct sales to the public. 2
FUNCTIONAL DESCRIPTION
2.1
Construction and devices
The digital weight indicators are fully described in Parts (Test) Certificate GB-1576. 2.2
Load cell
Any compatible load cell(s) may be used providing the following conditions are met: −
There is a respective OIML Certificate of Conformity (R60), or a Parts (Test) certificate (EN45501) issued for the load cell by a Notified Body responsible for type examination under Directive 2014/31/EU.
−
The certificate contains the load cell types and the necessary load cell data required for the manufacturer’s declaration of compatibility of modules (WELMEC 2, 2015, No 10), and any particular installation requirements. A load cell marked NH is allowed only if humidity testing to EN45501 has been conducted on this load cell.
−
The compatibility of the load cells and indicator is established by the manufacturer by means of the compatibility of modules calculation, contained in the above WELMEC 2 document, at the time of verification or declaration of EU conformity of type.
−
The load cell transmission must conform to one of the examples shown in the WELMEC Guide 2.4, “Guide for Load cells”.
2.2.2 When the requirements of 2.2.1 are met for systems comprising Avery T203 load cells (approved under R60/1991-GB-95.12) or Avery 8713 load cells (approved under TC2264), these load cells may be freely mixed within individual systems as these cells are technically compatible and freely interchangeable. 2.2.3 When the requirements of 2.2.1 are met for systems comprising Avery T204 load cells (approved under R60/1991-GB-97.02) or Avery 8708 load cells (approved under R60/1991-GB-94.01), these load cells may be freely mixed within individual systems as these cells are technically compatible and freely interchangeable. The mounting arrangement shall be such that the mechanical differences are taken into account. 2.2.4 When the requirements of 2.2.1 are met for systems comprising Avery T302 load cells (approved under R60/1991-GB-00.01) or Avery T302i load cells (approved under R60/2000-GB1-04.03) or Avery T302x load cells (approved under R60/2000-GB1-09.10), these load cells may be freely mixed within individual systems as these cells are technically compatible and freely interchangeable.
2/5
The mounting arrangement shall be such that the mechanical differences are taken into account. 2.2.5
A maximum of four load receptors may be connected to the indicator.
3
TECHNICAL DATA
3.1 GB-1576.
Technical data for the indicators is provided in Parts (Test) Certificate
4
PERIPHERAL DEVICES AND INTERFACES
4.1
Interfaces
The instrument may have the following interfaces type (fully described in GB-1576): − − − − − − − − − − − − − − − − − − − −
4.2
Load cell 4-wire or 6-wire shielded connections Logic level inputs Open collector outputs RS232/422/485 10/100 Ethernet USB Host Analogue outputs Current Loop Wireless LAN 802.11 b/g USB Device Bluetooth DeviceNet Profibus Power over Ethernet DC inputs DC outputs AC inputs AC outputs GSE & 1310 legacy interfaces Programmable Digital I/O Interfaces Peripheral devices
The following peripheral devices may be connected to the interfaces provided: −
Peripheral devices that have been issued with a Parts (Test) Certificate by a Notified Body responsible for type examination under Directive 2014/31/EU; or
−
Peripheral devices without a Parts (Test) Certificate under the following conditions: - it bears the CE marking for conformity to the EMC Directive; - it is not capable of transmitting any data or instruction into the weighing instrument, other than to release a printout, checking for correct data transmission or validation; - it prints weighing results and other data as received from the weighing instrument without any modification or further processing;
3/5
- it complies with the applicable requirements of EN45501, i.e. 4.2, 4.4, 4.6 and 4.7. A printing device may print additional information such as date or number to identify the printed weighing result(s) or sets of weighing results. 5
APPROVAL CONDITIONS
This certificate is issued subject to the following conditions: 5.1
Inscriptions
5.1.1
The instrument shall bear the following inscriptions on or near the display: − − −
5.1.2
Max Min e= The instrument shall bear the following inscriptions:
− − − − − − −
CE marking Supplementary metrology marking Identification number of the notified body Number of the EU-type examination certificate Manufacturer’s name, registered trade name or registered trade mark and postal address Accuracy class Type, batch or serial number
and, when applicable: − for instruments consisting of separate but associated units: identification mark on each unit; − scale interval if it is different from e, in the form d = H; − maximum additive tare effect, in the form T = + H; − maximum subtractive tare effect if it is different from Max, in the form T = – H; − tare interval if it is different from d, in the form dT = H; − maximum safe load if it is different from Max, in the form Lim H; − the special temperature limits, in the form H oC/H oC; − ratio between load receptor and load. The markings and inscriptions shall fulfil the requirements of Article 6, Article 16, Article 17 and Point 1 of Annex III of Directive 2014/31/EU. 6
LOCATION OF SEALS AND VERIFICATION MARKS
6.1 The data plate is secured, either by sealing or by being of a form such that it is destroyed when removed. 6.2 Components that may not be dismantled or adjusted by the user (jumper on main board when applicable, load cell connection) must be secured. Common serial numbers, a wire and seal solution or a tamper-evident sticker (bearing a securing mark) may be used. The securing mark may be either:
− −
a mark of the manufacturer and/or manufacturer’s representative, or an official mark of a verification officer.
4/5
6.3 When software sealing is used, the CONFIG and CAL counters’ values shall be written on a tamper-evident label on or near the rating plate. 7
ALTERNATIVES
7.1 The ZM505 / ZM510 / ZM605 / ZM615 indicators may be connected to a maximum of two Avery Weigh-Tronix BSQ series digital bases (approved under Test Certificate GB-1556) to form a complete single-interval or multi-range Class III NonAutomatic Weighing Instrument. The BSQ base(s) must be set up for operation in “Raw, unprocessed digital counts” mode (as described in section 2.3 of GB-1556). The instrument has the following characteristics for the BSQ weighing input(s): n ≤ 10,000 for Class III instruments with single interval ni ≤ 10,000 for Class III instruments with multi-range operation, per weighing range, with a maximum of three weighing ranges n ≤ 1,000 for Class IIII instruments with single interval 7.2 The ZM505 / ZM510 / ZM605 / ZM615 indicators may be connected to a load receptor fitted with a maximum of 8 Avery T301 digital load cells (approved under R60/1991GB-94.03) or Avery Weigh-Tronix T301x digital load cells (approved under R60/1991-GB111.05), with the following characteristics: Load Cell Avery T301 Avery T301x
Emax 30000 kg 30000 kg
Mounting Drawing 70714-342 AWT15-000226
VMin Emax/9375 Emax/9375
nmax 6000 6000
This configuration forms a complete single-interval, multi-interval or multi-range Class III or Class IIII Non-Automatic Weighing Instrument. These load cells may be freely mixed within individual systems as these cells are technically compatible and freely interchangeable. The T301 / T301x load cells are connected to a digital junction box, containing a load cell scanning PCB, either part no. 71014-875 or AWT25-001315. Any combination of scanning PCB and load cells listed in 7.2 is permitted. The instrument has the following characteristic: n ≤ 6,000 for Class III instruments with single interval ni ≤ 3,000 for Class III instruments with multi-range or multi-interval operation, per weighing range, with a maximum of three weighing ranges n ≤ 1,000 for Class IIII instruments with single interval 7.3 Having the indicator fitted with one or more additional load cell interfaces and connected to a maximum of four load receptors. The interfaces may be a combination of analogue load cells, BSQ platforms (see section 7.1), or T301/T301x digital load cells (see section 7.2) The primary display will indicate the active load receptor, as 1 / 2 / 3 / 4, and associated weight display. Selection of the load receptor is achieved by pressing the “Scale” key. 7.4 The ZM505 / ZM510 / ZM605 / ZM615 may be used for “direct sales to the public” if the pre-set tare device and “Select” function key are inactive.
5/5
8
ILLUSTRATIONS
Figure 1 Figure 2 Figure 3 Figure 4 Figure 5 Figure 6 Figure 7 Figure 8 Figure 9 Figure 10
ZM505-SD3 Indicator ZM505-SP3 Indicator ZM510-SD4 Indicator ZM510-SP4 Indicator ZM605-SD4 Indicator ZM605-SP4 Indicator ZM615-SD5 Indicator ZM615-SP5 Indicator Rating plate example (Mains Powered) Rating plate example (DC Powered)
CERTIFICATE HISTORY ISSUE NO.
DATE
DESCRIPTION
UK 3042
21 July 2016
Type-examination certificate first issued.
UK 3042 Revision 1
13 September 2017
Wording in 5.1.2 updated to better reflect 2014/31/EU. Addition of ZM505 / ZM605 / ZM615 series. “Programmable Digital I/O” added to 4.1. Renumbering and addition of figures to reflect additional models in the range.
6/5
Figure 1
ZM505-SD3 Indicator
Figure 2
ZM505-SP3 Indicator
7/11
Figure 3
ZM510-SD4 Indicator
Figure 4
ZM510-SP4 Indicator
8/11
Figure 5
ZM605-SD4 Indicator
Figure 6
ZM605-SP4 Indicator
9/11
Figure 7
ZM615-SD5 Indicator
Figure 8
ZM615-SP5 Indicator
10/11
Figure 9
Figure 10
Rating plate example (Mains Powered)
Rating plate example (DC Powered)
Crown copyright 2017 NMO Regulatory Delivery directorate Department for Business, Energy & Industrial Strategy This material may be freely reproduced except for sale