iMAG 4700 - Seametrics

Report 0 Downloads 137 Views
iMAG 4700

Mechanical Quick Start Guide Components

Power/Output Cable (DC configuration)

Cable Glands 15 Pin Connector (DC configuration)

Security Seal

Meter and Indicator

Optional

Protective faceplate cover and lanyard

Output Cable (Battery configuration)

Second Output Cable (DC configuration)

Recommended Tools Recommended:

Optional:

• T-15 Torx head driver

• Channel locks

• 0.098 inch (small) flat head screw driver • Crescent Wrench

253.872.0284 seametrics.com

iMAG 4700

Mechanical Quick Start Guide

Positioning Choose a position that will ensure a full pipe.

Choose a position that will minimize flow distortion. 5X

1X

2X

1X

5X

Recommended: Improved accuracy results from unimpeded electrodes

Not Ideal: Air bubbles and sediment on the electrodes can affect accuracy

Intermittent air bubbles miss electrode Electrode moved from top by rotating meter

1X

Intermittent air bubbles pass over electrode

Electrodes free from sediment build-up

Possible sediment build-up

Grounding Plastic Pipe

Metal Pipe Equalization Lug

Meter Flange Gasket

Grounding Ring Gaskets

Pipe Flange Equalization Lug

Grounding Ring Gaskets

Lockwasher Flange Plate

Plastic Pipe

Plastic Pipe

Ground Clamp

(Exothermically weld when corrosion is a concern)

Meter Flange Metal Pipe

Warnings

Gaskets

Pipe Flange

Earth #6, #8, or #12 AWG Stranded Copper Ground Wire < 5’

Metal Pipe

8’ Ground Rod

Refer to instruction manual for further details.

• This meter is sensitive to heat. Do not weld or flame cut within 10’ of meter. Heat will damage electronics and liner. (Heat will void warranty.) • It is highly recommended that full face gaskets be installed on each end of the meter to protect the liner. (Damage to liner will void warranty.)

• Tube must be completely aligned with adjoining flanges. • Bolts must be tightened evenly. Do not over tighten one side to fill a gap. • Improper sealing of cables or cable glands will void warranty. • Install security seal during installation if regulations require.

Refer to the Electrical Quick Start Guide or the complete instruction manual for details on wiring. Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA (P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com